20 pA range (0.1 fA resolution), Accuracy: ±(2.0 % of rdg. +30 dgt.) 200 pA range (1.0 fA resolution), Accuracy: ±(1.0 % of rdg. +30 dgt.) 2 nA range (10 fA resolution), Accuracy: ±(0.5 % of rdg. +20 dgt.) 20 nA range (100 fA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 200 nA range (1 pA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 2 μA range (10 pA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 20 μA range (100 pA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 200 μA range (1 nA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) *2 mA range (1 nA resolution), Accuracy: ±(0.5 % of rdg. +30 dgt.) (1) Measurement speed SLOW2 (internal integration time 13PLC) (2) At a temperature of 23 °C ±5 °C with humidity of 85% rh (3) 2 mA range (Measurement speed FAST only)
Resistance measurement capabilities
50 Ω to 2 × 10 ^19 Ω Note: Resistance measurement accuracy is defined by the current range accuracy and voltage setting accuracy.
Setting voltage range (Accuracy)
0.1 to 100.0 V, 100 mV resolution, Accuracy: ±0.1 % of setting ±0.05% f.s. 100.1 to 1000 V, 1 V resolution, Accuracy: ±0.1 % of setting ±0.05% f.s.
Current Limiter
0.1 to 250.0 V: 5/ 10/ 50 mA, 251 to 1000 V: 5/ 10 mA, to 2000 V:1.8 mA
USB driver) ×1, EXT I/O male connector ×1, Short plug ×1
Art.-No. SM-7120 : SUPER MEGOHM METER.
Number of channels
1 ch
DC current measurement
20 pA range (0.1 fA resolution), Accuracy: ±(2.0 % of rdg. +30 dgt.) 200 pA range (1.0 fA resolution), Accuracy: ±(1.0 % of rdg. +30 dgt.) 2 nA range (10 fA resolution), Accuracy: ±(0.5 % of rdg. +20 dgt.) 20 nA range (100 fA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 200 nA range (1 pA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 2 μA range (10 pA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 20 μA range (100 pA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 200 μA range (1 nA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) *2 mA range (1 nA resolution), Accuracy: ±(0.5 % of rdg. +30 dgt.) (1) Measurement speed SLOW2 (internal integration time 13PLC) (2) At a temperature of 23 °C ±5 °C with humidity of 85% rh (3) 2 mA range (Measurement speed FAST only)
Resistance measurement capabilities
50 Ω to 2 × 10 ^19 Ω Note: Resistance measurement accuracy is defined by the current range accuracy and voltage setting accuracy.
Setting voltage range (Accuracy)
0.1 to 100.0 V, 100 mV resolution, Accuracy: ±0.1 % of setting ±0.05% f.s. 100.1 to 1000 V, 1 V resolution, Accuracy: ±0.1 % of setting ±0.05% f.s.
[SM7120 only] 1000 to 2000 V,1 V resolution, Accuracy: ±0.2 % of setting ±0.10% f.s.
Current Limiter
0.1 to 250.0 V: 5/ 10/ 50 mA, 251 to 1000 V: 5/ 10 mA, to 2000 V:1.8 mA
USB driver) ×1, EXT I/O male connector ×1, Short plug ×1
Art.-No. SM-7420 : SUPER MEGOHM METER.
Number of channels
4ch
DC current measurement
20 pA range (0.1 fA resolution), Accuracy: ±(2.0 % of rdg. +30 dgt.) 200 pA range (1.0 fA resolution), Accuracy: ±(1.0 % of rdg. +30 dgt.) 2 nA range (10 fA resolution), Accuracy: ±(0.5 % of rdg. +20 dgt.) 20 nA range (100 fA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 200 nA range (1 pA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 2 μA range (10 pA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 20 μA range (100 pA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 200 μA range (1 nA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) *2 mA range (1 nA resolution), Accuracy: ±(0.5 % of rdg. +30 dgt.) (1) Measurement speed SLOW2 (internal integration time 13PLC) (2) At a temperature of 23 °C ±5 °C with humidity of 85% rh (3) 2 mA range (Measurement speed FAST only)
Resistance measurement capabilities
50 Ω to 2 × 10 ^19 Ω Note: Resistance measurement accuracy is defined by the current range accuracy and voltage setting accuracy.