Portable Spectroradiometers
ILT950 / ILT950-UV
모델명 ILT950 / ILT950-UV
제조사 ILT INC
원산지 미국
분류명 Spectrometers

Specifications


Two Versions :

  • ILT950 : 200-1100 nm
    Resolution : 1.4 nm with 25 micron slit**
  • ILT950-UV : 200-450 nm
    Resolution : >1 nm with 50 micron slit**

Radiometric Accuracy :

  • 200nm - 350nm : ±20%**
  • >350nm - 400 nm : ±10%**
  • >400nm - 900nm : ±5%
  • >900nm - 1050nm:  ±10%
NIST-Traceable/ISO17025 Accredited Calibration**Data Transfer Speed : 2ms/scan
Detector: CMOS Linear Image Sensor16 bits, 2 MHz
Focal Length : 75 mmTrigger Compatible
BlazeSynchronization Capability
Dynamic Range : 3300Temperature Range: 15 - 40°C
Symmetrical Czerny-TurnerSize : 1-3/5” H x 5” W x 7” L
Stray light : <0.3%Dynamic Dark Correction : Yes
Signal/Noise : 300:1Non-Linearity Calibration : Yes
Integration Time : 30 µs - 59s Wavelength Accuracy : ±.5 nm

** (Note : Dual range calibration with D2 and QTH light source is strongly recommended for broadband measurements encompassing

UV below 350 nm due to low output of the QTH lamp which increases the difficulty of accurate UV calibration.

The ILT950 can be calibrated with QTH light source only if UVB/C accuracy is not consequential.)


파일명 확장자 다운로드
 ILT950_ILT950-UV[Brochure].pdf (979KB)
 ILT SERIES Spectrometer Comparison Guide.pdf (345KB)
 Optional Accessories[Brochure].pdf (247KB)


상호 : 주식회사 마스  |  (우)13209 경기도 성남시 중원구 사기막골로 45번길 14 우림2차 A-1610호   |   Tel : 031-735-6310 / 6320   |   Fax : 031-735-6334
대표이사 / 개인정보관리책임자 : 안승만   |   사업자등록번호 : 129-86-80575   |   통신판매업등록번호 : 제2013-경기성남-0854호   |   메일 : mmm@m-m.co.kr

COPYRIGHT© MAS CO.,LTD. ALL RIGHTS RESERVED.