General |
|
Memory |
32,768 readings
(all channels enabled) |
Reading Rate |
1 reading every second up to 1
reading every 24 hours |
Battery Life |
2 years typical
at 25 °C, 15 minute reading intervals |
Material |
316 Stainless Steel/Radel |
Dimensions |
1.7 in x 0.97 in
x 0.97 in (42 mm x 24.6 mm x 24.6 mm) |
Operating
Environment |
-40 °C to +80 °C (-40 °F to
+176 °F), 0 %RH to 100 %RH (non-condensing) |
Required
Interface Package |
IFC400 or IFC406 |
Approvals |
CE |
Temperature |
|
Temperature Sensor |
Resistance
Temperature Detector (RTD) |
Temperature Range |
-40 °C to +80 °C (-40 °F to
+176 °F) |
Temperature
Resolution |
0.01 °C (0.018
°F) |
Temperature
Response |
t63 - 8 minutes to 63% of
change |
Time in Air |
t90
- 68 minutes to 90% of change |
Temperature
Calibrated Accuracy |
±0.5 °C (0 °C to
+55 °C) |
Humidity |
|
Humidity Sensor |
Capacitive
Polymer |
Humidity Range |
0 %RH to 100 %RH
(non-condensing) |
Humidity
Resolution |
0.1 %RH |
Humidity Response
Time in Air |
t90 - 2 minutes to 90% of
change |
Humidity
Calibrated Accuracy |
±3 %RH maximum |
Humidity
Specified Accuracy Range |
25 %RH to 75 %RH |
+20 °C to +40 °C |
Hysteresis Error 1 % Typical, 3% Maximum.
|